CPSC2012-07-18

TRUMPF Medical Systems Recalls Television Mounting System Due to Falling Hazard

Ondal Medical Systems of America Inc., of Sandston, VA
Hazard

The arm joints on the television mounting system do not have a washer at the top joint which can result in excessive wear of the stop pins on the arm. This can allow the arm system and items connected to the arm to fall and injure the user or bystanders.

Remedy
Secure the unit to prevent movement and contact TRUMPF Medical Systems for information on receiving a repair kit that includes the missing washers.
Sold at
TRUMPF Medical Systems Inc. nationwide From January 2005 through December 2011 for about $2,100.
Affected count
215 distributed to 5 hospitals
Injuries reported
The firm received one report of a mounting system falling while being positioned. No injuries have been reported.
Manufactured in
Germany
Remedy type
Repair
Products
Ondal AC2000 (PDI) Television Mounting System
Consumer contact
For additional information, contact TRUMPF Medical Systems Inc. toll-free at (888) 474-9359 between 8 a.m. and 5 p.m. ET Monday through Friday, or email Lindsey.ronnenberg@us.trumpf.com

Official notice

https://www.cpsc.gov/Recalls/2012/TRUMPF-Medical-Systems-Recalls-Television-Mounting-System-Due-to-Falling-Hazard