FDA-Device2025-10-08Class IIPROCESSING DEFECT
Philips Allura Xper FD10C medical imaging system recalled for BIOS battery depletion
BATTERYNationwide distribution
Contact Philips about Allura Xper FD10C system failure
The BIOS battery in some Philips Allura Xper FD10C X-ray systems may run out of power faster than expected. When this happens, the system will not start up, and there are no warning messages to alert users before the battery fails.
- If you operate an Allura Xper FD10C system, contact Philips Medical Systems immediately for guidance
- Do not attempt to restart the system repeatedly if it fails to start
- Follow Philips's instructions for battery replacement or service
Hazard
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
Sold states
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Affected count
2 units (US only)
Manufactured in
Veenpluis 4-6, N/A, Best, N/A, Netherlands
Products
Allura Xper FD10C; Model Number: 722001; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0001-2026More PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls
- FDA-Device2026-06-03Philips Azurion imaging systems recalled for hard drive failure risk affecting imaging functionality
- FDA-Device2026-06-03Philips Allura medical imaging systems recalled for hard drive failure risk
- FDA-Device2026-05-27Philips Azurion X-ray systems recalled for cooling hose degradation risk
- FDA-Device2026-05-27Philips Integris-Allura X-ray systems recalled for cooling unit oil leakage risk
- FDA-Device2026-05-27Philips Azurion X-ray systems recalled for impaired table movement due to control wear
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief