FDA-Device2025-10-08Class IIPROCESSING DEFECT

Philips Allura Xper FD10 X-ray system recalled for rapid BIOS battery depletion

BATTERYNationwide distributionMedical devices

What this means

Certain Philips Allura Xper FD10 X-ray systems may have a BIOS battery that depletes faster than expected, preventing the system from starting up. The system will not display any warning before the battery runs out of power.

  • Contact Philips if you operate one of the affected model numbers (722003, 722010, or 722026)
  • Check your device serial number against the FDA recall list
  • Follow Philips guidance on battery replacement or system service
Get recalls like this in a free plain-English brief every Sunday.Get alerts ↓
The agency's original wording
Hazard

The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.

Intended for
Adults
Illnesses reported
None reported
Sold states
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Affected count
2317 units (497 US, 1820 OUS)
Manufactured in
Veenpluis 4-6, N/A, Best, N/A, Netherlands
Products
Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0002-2026