FDA-Device2025-10-08Class IIPROCESSING DEFECT
Philips Allura Xper FD10 X-ray system recalled for rapid BIOS battery depletion
BATTERYNationwide distribution
Check your Philips Allura Xper FD10 system
Certain Philips Allura Xper FD10 X-ray systems may have a BIOS battery that depletes faster than expected, preventing the system from starting up. The system will not display any warning before the battery runs out of power.
- Contact Philips if you operate one of the affected model numbers (722003, 722010, or 722026)
- Check your device serial number against the FDA recall list
- Follow Philips guidance on battery replacement or system service
Hazard
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
Sold states
Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
Affected count
2317 units (497 US, 1820 OUS)
Manufactured in
Veenpluis 4-6, N/A, Best, N/A, Netherlands
Products
Allura Xper FD10; Model Numbers: (1) 722003, (2) 722010, (3) 722026; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0002-2026More PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls
- FDA-Device2026-06-03Philips Azurion imaging systems recalled for hard drive failure risk affecting imaging functionality
- FDA-Device2026-06-03Philips Allura medical imaging systems recalled for hard drive failure risk
- FDA-Device2026-05-27Philips Azurion X-ray systems recalled for cooling hose degradation risk
- FDA-Device2026-05-27Philips Integris-Allura X-ray systems recalled for cooling unit oil leakage risk
- FDA-Device2026-05-27Philips Azurion X-ray systems recalled for impaired table movement due to control wear
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief