FDA-Device2024-10-09Class II

IntelliVue G7m Anesthesia Gas Module, Product Number 866173

Philips North America
Hazard

The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, INOP (inoperable) alarm. The pump may fail prematurely when there is friction caused by an abrasion. This recall is for an IMPORTANT PRODUCT NOTICE dated 1/18/23.

Sold states
US Nationwide. Global Distribution.
Affected count
8,449 devices
Manufactured in
222 Jacobs St, N/A, Cambridge, MA, United States
Products
IntelliVue G7m Anesthesia Gas Module, Product Number 866173

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0007-2025

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.