FDA-Device2024-10-09Class II

Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility.

Outset Medical, Inc.
Hazard

Blood leak sensor, a component of hemodialysis device, may show decreased sensitivity, at the lowest level, specified as 0.35 mL/min blood leak rate, which could delay triggering the ALARM_DIALYZER_BLOOD_LEAK alarm, if blood loss, less-than 300ml, occurs, the patient may experience hypotension, fatigue, cramps, dizziness, headaches, nausea, and shortness of breath.

Sold states
US: CA, NY, IL, MO, KY, AZ, PA, FL, GA, AK, NM, VA, NC, SC, OK, AL, MA, OH, MI, WA, ID, TX, TN, WY, IA, LA, NE, IN, AR, NV, WV, CO, NJ, MD, KS, UT, CT, MN, MS, WI, HI, DE, VT, MT, OR, ND. OUS: UAE
Affected count
2,422
Manufactured in
3052 Orchard Dr, N/A, San Jose, CA, United States
Products
Tablo Console, REF: PN-0003000, PN-0006000, part of the Tablo Hemodialysis System. The Tablo Hemodialysis System is indicated for use in patients with acute and/or chronic renal failure, with or without ultrafiltration, in an acute or chronic care facility.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0019-2025