FDA-Device2021-10-13Class II

TRULIANT Knee System including TRULIANT Tibial Inserts, also labeled as the following: a. TRULIANT CR Tibial Inserts; b. TRULIANT CR Slope + Tibial Inserts; c. TRULIANT CR Slope ++ Tibial Inserts; d. TRULIANT CRC Tibial Inserts; e. TRULIANT PS Tibial Inserts; f. TRULIANT PSC Tibial Inserts. NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022.

Exactech, Inc.
Hazard

Inserts were packaged in vacuum bags that lacked an additional oxygen barrier layer.

Sold states
Worldwide distribution - US Nationwide distribution and the countries of ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BRAZIL, CANADA, CHILE, CHINA, COLOMBIA, CYPRUS, ECUADOR, FRANCE, FRENCH POLYNESIA, GERMANY, GHANA, GREAT BRITAIN, GREECE, GUATEMALA, INDIA, ITALY, JAPAN, JORDAN, KOREA, REPUBLIC OF, LEBANON, LUXEMBOURG, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, POLAND, PORTUGAL, ROMANIA, SINGAPORE, SPAIN, SWITZERLAND, SYRIAN ARAB REPUBLIC, THAILAND, TUNISIA, TURKEY, VENEZUELA .
Affected count
31,201 devices
Manufactured in
2320 Nw 66th Ct, Gainesville, FL, United States
Products
TRULIANT Knee System including TRULIANT Tibial Inserts, also labeled as the following: a. TRULIANT CR Tibial Inserts; b. TRULIANT CR Slope + Tibial Inserts; c. TRULIANT CR Slope ++ Tibial Inserts; d. TRULIANT CRC Tibial Inserts; e. TRULIANT PS Tibial Inserts; f. TRULIANT PSC Tibial Inserts. NOTE: Information reflects Exactech recall expansion initiated on 02/07/2022.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0023-2022