FDA-Device2012-10-17Class II

Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL) Product Usage: Enzyme n Immunoassay, Phenobarbital

Siemens Healthcare Diagnostics, Inc.
Hazard

Firm has confirmed highter imprecision with the recalled lots on Beckman Coulter AU(R) Systems. Also, if the Syva(R) Emit(R) 2000 Phenobartbital Assay (4D019UL) lots D4, D5, D6, and/or D7 as an application on AU(R) Systems is used, the issue may be experienced. Firm has found that within-run imprecison of the recalled lots when run on an AU(R) System may be higher that what would normally be obs

Sold states
United States Nationwide Distribution in the states of: CA, FL, KS, MA, MN, MS, NM, NY, NC, OR, PA, TN, TX, UT, VA, and WA.
Affected count
79,744 cartons
Manufactured in
500 GBC Drive, Mailstop 514, PO BOX 6101, Newark, DE, United States
Products
Emit(R) 2000 Phenobarbital Assay (OSR4D229) and Syva(R) Emit(R) Phenobarbital Assay (4D019UL) Product Usage: Enzyme n Immunoassay, Phenobarbital

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0030-2013

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief