FDA-Device2024-10-16Class II

Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390, Software Revision 4.0.1

Philips North America
Hazard

During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.

Sold states
Worldwide - US Nationwide distribution.
Affected count
591 devices
Manufactured in
222 Jacobs St, Cambridge, MA, United States
Products
Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390, Software Revision 4.0.1

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0035-2025