FDA-Device2013-11-06Class II

Xoran MiniCAT00202

Xoran Technologies, Inc.
Hazard

The MiniCAT CT Scanners subject to this notification failed to meet Xoran's specifications for dose rate of 125 ¿ 3 kVp. These 18 complaints received include dose rates both on the high and low end of Xorans specifications.

Sold states
Worldwide Distribution - Nationwide
Affected count
320
Manufactured in
5210 S State Rd, N/A, Ann Arbor, MI, United States
Products
Xoran MiniCAT00202

Official notice