FDA-Device2023-10-11Class II

Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299

Blue Belt Technologies, Inc
Hazard

CORI software was missing a planning stage that appears in cases of significant knee deformation (outside of the range of 3 degrees valgus to 7 degrees varus) when the surgeon chooses to define the femoral rotational reference via the posterior condylar axis

Sold states
US Nationwide distribution.
Affected count
56 units
Manufactured in
2875 Railroad St, N/A, Pittsburgh, PA, United States
Products
Real Intelligence CORI RI.KNEE v2.0 TKA (Total Knee Arthroplasty) software version 2.0 (2.0.0.5) Product Number: ROB10299

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0041-2024