FDA-Device2025-10-15Class II

FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33

FUJIFILM Healthcare Americas Corporation
Hazard

It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting.

Sold states
Domestic: AK, CA, FL, IL, MA, NH, OH, WI.
Affected count
16 units
Manufactured in
81 Hartwell Ave Ste 300, Lexington, MA, United States
Products
FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0044-2026

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief