FDA-Device2012-10-17Class II

IDS-iSYS Intact PINP Control Set, Catalog Number IS-4030. Packaged in cardboard Box containing plastic bottles. The IDS-iSYS Intact amino-terminal propeptide of type I procollagen (Intact PINP) Control Set is used for quality control of the IDS-iSYS Intact PINP Assay on the IDS-iSYS Multi-Discipline Automated Analyser.

Immunodiagnostics Systems Ltd
Hazard

It has been noted by staff at IDS Inc. that stock of the affected products held at IDS Inc. are labeled with IVD labels although they are not registered for sale as IVDs in the United States. The products should have been labeled as Research Use Only or Investigational use only dependent on their usage.

Sold states
Nationwide Distribution including NY, CA, ME and CT.
Affected count
29
Manufactured in
10 Didcot Way Bolden Business Park, N/A, Boldon Tyne & Wear, N/A, United Kingdom
Products
IDS-iSYS Intact PINP Control Set, Catalog Number IS-4030. Packaged in cardboard Box containing plastic bottles. The IDS-iSYS Intact amino-terminal propeptide of type I procollagen (Intact PINP) Control Set is used for quality control of the IDS-iSYS Intact PINP Assay on the IDS-iSYS Multi-Discipline Automated Analyser.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0045-2013