FDA-Device2017-11-08Class II

THERMOCOOL SMARTTOUCH SF Uni-Directional Product Codes D-1347-01-S, D-1347-02-S, D-1347-03-S

Biosense Webster, Inc.
Hazard

Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.

Sold states
US, Japan, Canada, Australia, New Zealand, Brazil, Croatia, Korea, Austria, Colombia, Italy, Portugal, Argentina, South Africa, Belgium, France, Germany, Ireland, Poland, Netherlands, Sweden, Denmark, Finland, Switzerland, Great Britain, Hungary, Czech Republic, China
Affected count
357 units
Manufactured in
15715 Arrow Hwy, N/A, Irwindale, CA, United States
Products
THERMOCOOL SMARTTOUCH SF Uni-Directional Product Codes D-1347-01-S, D-1347-02-S, D-1347-03-S

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0054-2018