FDA-Device2016-10-19Class II

The Medix NatalCare LX Intensive Care Incubator. Neonatal incubator. Item numbers: DMNCLX1; KITNC-STLX0003; KITNC-STLX0006; and KITNC-STLX0007.

Natus Medical Incorporated
Hazard

NatalCare LX incubators may have a defective base where the weld holding the wheel bushing to the tube is inadequate. There were several complaints of wheels bending out of alignment.

Sold states
US Distribution to the states of : AL, CA, FL, IL, IN, KY, LA, MA, MS, NJ, NY, OK, PA, SC and VA, and Internationally to Pakistan.
Affected count
104 units
Manufactured in
5900 1st Ave S, N/A, Seattle, WA, United States
Products
The Medix NatalCare LX Intensive Care Incubator. Neonatal incubator. Item numbers: DMNCLX1; KITNC-STLX0003; KITNC-STLX0006; and KITNC-STLX0007.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0056-2017