FDA-Device2017-11-08Class II

THERMOCOOL SF NAV Catheters Model Numbers - BNI35FJCT, BNI35DFCT, D-131503-S, D-131504-S

Biosense Webster, Inc.
Hazard

Biosense Webster, Inc. has recently received an increased number of complaints related to the display of Alert 402 on the CARTO 3 System for certain lots of THERMOCOOL brand catheters. Alert 402 implies a "Map: magnetic distortion" when connected to CARTO 3 System. This issue may subsequently lead the physician to ablate in an unintended area when delivering RF energy.

Sold states
US, Japan, Canada, Australia, New Zealand, Brazil, Croatia, Korea, Austria, Colombia, Italy, Portugal, Argentina, South Africa, Belgium, France, Germany, Ireland, Poland, Netherlands, Sweden, Denmark, Finland, Switzerland, Great Britain, Hungary, Czech Republic, China
Affected count
329 units
Manufactured in
15715 Arrow Hwy, N/A, Irwindale, CA, United States
Products
THERMOCOOL SF NAV Catheters Model Numbers - BNI35FJCT, BNI35DFCT, D-131503-S, D-131504-S

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0058-2018

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief