FDA-Device2024-10-23Class II
1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and smoothing of bone, bone cement and teeth in a variety of surgical procedures including but not limited to dental, ENT (ear, nose and throat), neuro, spine and endoscopic applications Catalog Number: 5820-066-012
Stryker Corporation
Hazard
May exhibit endotoxin levels greater than 2.15 EU/device, fever and inflammation may occur from endotoxin exposure and meningitis may only occur if there is contact with cerebrospinal fluid and thus, neurosurgical/spinal procedures may be considered a population at greatest risk.
Sold states
Worldwide distribution - US Nationwide and the countries of Canada, Columbia.
Affected count
307 units
Manufactured in
1941 Stryker Way, Portage, MI, United States
Products
1.2mm Cross Cut Fissure - Intended for use in the cutting, drilling, reaming, decorticating, shaping and smoothing of bone, bone cement and teeth in a variety of surgical procedures including but not limited to dental, ENT (ear, nose and throat), neuro, spine and endoscopic applications Catalog Number: 5820-066-012
Official notice