FDA-Device2019-10-16Class II

BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 662877, when using BD Trucount Tubes

Becton, Dickinson and Company, BD Biosciences
Hazard

Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.

Sold states
U.S.: CA, KS, NJ, MI, NY, PA, IL, IN, VA, WA, MD, IA, OK, RI, FL, CO, OH, TX, TN, GA, MO, MN, HI, VT, MA, NE. O.U.S. (Foreign): Germany, Italy, South Korea, Japan, United Kingdom, Poland, France, Belgium, Canada, Latvia, India, Taiwan, Ireland, Algeria, Israel, Iraq, Netherlands, Malaysia, Australia, Sweden, Colombia, Switzerland, Slovakia, Spain, Brazil, Austria, Peru, Iran, Croatia, Lithuania, Czech Republic, Saudi Arabia, Finland, Qatar, South Africa, Denmark, Turkey, Mexico, Indonesia, Ecuador, Vietnam, Singapore
Affected count
8
Manufactured in
2350 Qume Dr, N/A, San Jose, CA, United States
Products
BD FACSLyric Flow Cytometer, 3-Laser 8 Color Instrument, REF: 662877, when using BD Trucount Tubes

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0068-2020