FDA-Device2016-10-19Class II

OEC Brivo 715 Prime, OEC Brivo 785 Essential, OEC Brivo 865 Advance Product Usage: The OEC Brivo Mobile C-Arm X-Ray Products are designed to provide digital spot film imaging and fluoroscopic image guidance for all adult and pediatric populations for orientations between patient anatomy and surgical instruments. The product is used for general surgical applications and musculoskeletal procedures to visualize, for example, implant localizations or needle positions for aspirations, injections

GE OEC Medical Systems, Inc
Hazard

GE Healthcare Surgery announces a voluntary field action for the OEC Brivo 715 Prime, OEC Brivo 685 Essential, and OEC Brivo 865 Advance because the circuit board that controls vertical motion of the C-arm may fail, resulting in the C-Arm moving up or down without command.

Sold states
Worldwide Distribution - US Nationwide in the states of AL, AZ, CA, FL, KS, MS, NJ, NV, NY, PA, TX, UT, and WV. No Candaian or Va/govt/military consignees. There are 133 other foreign consignees.
Affected count
148 total
Manufactured in
384 N Wright Brothers Dr, N/A, Salt Lake City, UT, United States
Products
OEC Brivo 715 Prime, OEC Brivo 785 Essential, OEC Brivo 865 Advance Product Usage: The OEC Brivo Mobile C-Arm X-Ray Products are designed to provide digital spot film imaging and fluoroscopic image guidance for all adult and pediatric populations for orientations between patient anatomy and surgical instruments. The product is used for general surgical applications and musculoskeletal procedures to visualize, for example, implant localizations or needle positions for aspirations, injections or biopsy. The OEC Brivo is not indicated for interventional use.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0071-2017