FDA-Device2017-11-29Class I
Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device i
Spacelabs Healthcare, Ltd.
Hazard
Arkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen without audible or visible alarms.
Sold states
Nationwide Distribution to the states of NM, WY, FL, NC, MS, CO, CT, ME, and AL.
Affected count
110 units
Manufactured in
Foxholes Centre, Unit B, John Tate Road, Hertford, N/A, United Kingdom
Products
Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0072-2018Don't want to check this manually?