FDA-Device2020-10-14Class II

DISCOVERY ULTRA Advanced Staining System (RUO-Research Use Only) Material Number: 05987750001

Roche Diagnostics Operations, Inc.
Hazard

Potential for Fluid leaks within the instruments, creating an electrical short inside the power socket, which generates heat, smoke, and eventually burning and melting of the electrical cord connector and socket

Sold states
Nationwide Foreign: Canada, Austria, Belgium, Bulgaria, Croatia, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, UK, Algeria, Argentina, Aruba, Australia, Bahrain, Bosnia-Herz., Brazil, Chile, China, Colombia, Costa Rica, Cuba, Dominican Republic, Ecuador, Egypt, El Salvador, Hong Kong, India, Israel, Japan, Jordan, Kazakhstan, Kuwait, Malaysia, Martinque, Mexico, Monaco, Morocco, Namibia, New Caledonia, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Russia, Saudi Arabia, Singapore, S. Africa, S. Korea, Taiwan, Thailand, Ukraine, UAE, Vietnam
Affected count
257 (US); 208 (ex-US)
Manufactured in
9115 Hague Rd, N/A, Indianapolis, IN, United States
Products
DISCOVERY ULTRA Advanced Staining System (RUO-Research Use Only) Material Number: 05987750001

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0086-2021