FDA-Device2013-11-06Class II
chromID Salmonella Agar, Reference No. 43621. A selective isolation and differentiation medium for the detection of Salmonella in human specimens.
bioMerieux, Inc.
Hazard
Complaints for Salmonella strains growing on this medium producing an uncharacteristic white colored colony instead of the expected pale pink to mauve colored colonies. This could potentially lead to Salmonella colonies not being detected.
Sold states
Nationwide Distribution.
Affected count
90
Manufactured in
100 Rodolphe St, N/A, Durham, NC, United States
Products
chromID Salmonella Agar, Reference No. 43621. A selective isolation and differentiation medium for the detection of Salmonella in human specimens.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0087-2014Don't want to check this manually?