FDA-Device2016-10-26Class II

SDMS (SenTec Digital Monitoring System)

SenTec AG
Hazard

Sentec AG found that with one batch of an internal SenTec Monitor component (Docking Station Module) there is a risk that a fastening screw could become loose under certain circumstances. A loose screw within the monitor can create uncontrolled electrical connections, including short circuits, with the risk of smolder or fire and electrical shock (patient and/or operator).

Sold states
Worldwide Distribution - US Distribution to the states of : LA, MI, MO, OH, TN. and to the countries of : United Kingdom, The Netherlands and Australia.
Affected count
75 (54 US; 21 =OUS)
Manufactured in
Ringstrasse 39, N/A, Therwil, N/A, Switzerland
Products
SDMS (SenTec Digital Monitoring System)

Official notice