FDA-Device2016-10-26Class II

VITROS Chemistry Products LAC Slides, REF/Product Codes 843 3880 (300 slides per sales unit) and 815 0112 (90 slides per sales unit), Unique Device Identifier Nos. 10758750004911 and 10758750004454, Rx ONLY, IVD --- For in vitro diagnostic use only. VITROS Chemistry Products LAC Slides quantitatively measure lactate (LAC) concentration in plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Ortho-Clinical Diagnostics
Hazard

Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.

Sold states
Worldwide Distribution - US (Nationwide) Puerto Rico and Bermuda), Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, and Venezuela.
Affected count
US: 32,081 sales units; Foreign: 29,932 sales units
Manufactured in
100 Indigo Creek Dr, N/A, Rochester, NY, United States
Products
VITROS Chemistry Products LAC Slides, REF/Product Codes 843 3880 (300 slides per sales unit) and 815 0112 (90 slides per sales unit), Unique Device Identifier Nos. 10758750004911 and 10758750004454, Rx ONLY, IVD --- For in vitro diagnostic use only. VITROS Chemistry Products LAC Slides quantitatively measure lactate (LAC) concentration in plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0090-2017