FDA-Device2016-11-02Class II
URISYS 2400 Urine Analyzer Automated Urinalysis System Product Usage: The RSD 800 is a computer-controlled stand-alone sorting system for sorting barcoded samples. The system has an online connection to the host.
Roche Diagnostics Operations, Inc.
Hazard
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Sold states
US Nationwide Distribution
Affected count
9014 in total
Manufactured in
9115 Hague Rd, Indianapolis, IN, United States
Products
URISYS 2400 Urine Analyzer Automated Urinalysis System Product Usage: The RSD 800 is a computer-controlled stand-alone sorting system for sorting barcoded samples. The system has an online connection to the host.