FDA-Device2020-10-21Class III

Tina-quant Complement C4 ver.2, Catalog 05991994190

Roche Diagnostics Operations, Inc.
Hazard

Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.

Sold states
Domestic distribution nationwide.
Affected count
180 units
Manufactured in
9115 Hague Rd, N/A, Indianapolis, IN, United States
Products
Tina-quant Complement C4 ver.2, Catalog 05991994190

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0103-2021