FDA-Device2013-11-06Class II
Siemens RAPIDLab¿ 1240 system blood gas analyzer Product Usage: RAPIDLab 1200 Systems: The Rapidlab 1200 systems are intended for in vitro diagnostic use by healthcare professionals in the quantitative testing of human whole blood. The systems can determine the following parameters: RAPIDLab 1240 pH, pC02, p02 RAPIDLab 1245 pH, pC02, p02, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb RAPIDLab 1260 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate RAPIDLab 1265 pH, pC02, p02, Na+, K+, Ca++, Cl-,
Siemens Healthcare Diagnostics Inc
Hazard
RAPIDPoint¿ 400 Series, RAPIDPoint¿ 500, or RAPIDLab¿ 1200 Series blood gas analyzer, users can inadvertently enter a wrong patient ID on the demographics screen, which ultimately could lead to a misidentification of the sample
Sold states
Worldwide Distribution - USA Nationwide and in the countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia Herzeg., Botswana, Brazil, Brunei, Bulgaria, Chile, Colombia, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Fren.Polynesia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Kazakhstan, Kenya, Kuwait, Latvia, Lithuania, Luxembourg, Macedonia, Malaysia, Mayotte, Netherlands, New Caledonia, New Zealand, Norway, P.R. China, Philippines, Poland, Portugal, Qatar, Rep. of Yemen, Republic Korea, Romania, Russian Fed., Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, Turkmenistan, U.A.E., United Kingdom, Vatikancity, Venezuela, Vietnam,
Affected count
179 units
Manufactured in
2 Edgewater Dr, N/A, Norwood, MA, United States
Products
Siemens RAPIDLab¿ 1240 system blood gas analyzer Product Usage: RAPIDLab 1200 Systems: The Rapidlab 1200 systems are intended for in vitro diagnostic use by healthcare professionals in the quantitative testing of human whole blood. The systems can determine the following parameters: RAPIDLab 1240 pH, pC02, p02 RAPIDLab 1245 pH, pC02, p02, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb RAPIDLab 1260 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate RAPIDLab 1265 pH, pC02, p02, Na+, K+, Ca++, Cl-, glucose, lactate, tHb, nBili, F02Hb, FCOHb, FMetHb, FHHb
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0105-2014Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief