FDA-Device2016-11-02Class II
cobas p 471 centrifuge unit calculator/data processing module, for clinical use Product Usage: The cobas p312 pre-analytical system is a computer controlled stand-alone system for sorting and decapping sample tubes. As an option, the system works with an online connection.
Roche Diagnostics Operations, Inc.
Hazard
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Sold states
US Nationwide Distribution
Affected count
9014 in total
Manufactured in
9115 Hague Rd, Indianapolis, IN, United States
Products
cobas p 471 centrifuge unit calculator/data processing module, for clinical use Product Usage: The cobas p312 pre-analytical system is a computer controlled stand-alone system for sorting and decapping sample tubes. As an option, the system works with an online connection.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0108-2017Don't want to check this manually?