FDA-Device2021-10-20Class II

Evita V500 Ventilator with Installed CO2 Measurement Option-Intended for the ventilation of adult,pediatric and neonatal patients. Catalog Number: 8416400

Draeger Medical, Inc.
Hazard

Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds

Sold states
US Nationwide
Affected count
6,797 units
Manufactured in
3135 Quarry Rd, N/A, Telford, PA, United States
Products
Evita V500 Ventilator with Installed CO2 Measurement Option-Intended for the ventilation of adult,pediatric and neonatal patients. Catalog Number: 8416400

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0108-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief