FDA-Device2016-11-02Class II
cobas p 312 Pre-Analytical System calculator/data processing module, for clinical use Product Usage: The cobas p471 system is an automatic centrifuge module for centrifuging primary sample tubes, which will then be transported to the output module.
Roche Diagnostics Operations, Inc.
Hazard
Roche has received four reports of injury caused by the sharp edge of the center guide rail of an analyzers sample rack tray. The cleaning procedure for the sample rack tray has been revised to clarify the cleaning process.
Sold states
US Nationwide Distribution
Affected count
9014 in total
Manufactured in
9115 Hague Rd, Indianapolis, IN, United States
Products
cobas p 312 Pre-Analytical System calculator/data processing module, for clinical use Product Usage: The cobas p471 system is an automatic centrifuge module for centrifuging primary sample tubes, which will then be transported to the output module.