FDA-Device2021-10-20Class II
Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight. Catalog Number: 8417400
Hazard
Software 2.51 and Lower with Installed CO2 Measurement Option, may result in Restart of the Ventilator, brief cessation of ventilation can occur and loss of PEEP for approx. 8 seconds
Sold states
US Nationwide
Affected count
2,169 units
Manufactured in
3135 Quarry Rd, N/A, Telford, PA, United States
Products
Babylog VN500 Ventilator with Installed CO2 Measurement Option-intended for the ventilation of neonatal patients from 0.4 kg (0.88 lbs) up to 10 kg (22 lbs) and pediatric patients from 5 kg (11 lbs) up to 20 kg (44 lbs) bodyweight. Catalog Number: 8417400
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0109-2022Don't want to check this manually?