FDA-Device2014-10-29Class II

Mortara Surveyor S12 and Surveyor S19 Patient Monitors. Product Usage: They are indicated for use in adult & pediatric patient populations. They facilitate the monitoring of: Non-invasive blood pressure (NIBP), Impedance respiration, Invasive blood pressure (P1-P4), Temperature, Functional arterial oxygen saturation (SpO2), End-tidal & inspired CO2, ECG monitoring with arrhythmia and ST-segment analysis, 12-Lead resting ECG, Cardiac output (C)). The Mortara Surveyor Patient Monitor i

Mortara Instrument, Inc
Hazard

Martara Instrument, Inc. has recently became aware of a potential safety hazard involving our Surveyor rand S12 S19 Patient Monitors. A hazardous situation may be created when mounting these monitors using the 75mm VESA mounting hole pattern provided in the battery cover of these monitors. If the mounting screws provided by Mortara are not used and if the mounting screws used are longer than the p

Sold states
Worldwide Distribution: US (Nationwide) in the states of: AZ, AR, CA, DC, FL, ID, IL IN, KY, LA, ME, MD, MA, MI, MN, MO, MT, NV, NJ, NY, NC, OK PA, SC, TN, TX, VA, and WI; and internationally to: AUSTRALIA, BANGLADESH, CANADA, CHILE, CROATIA, EGYPT, GERMANY, ITALY, JORDAN, LEBANON, MALAYSIA, MOROCCO, OMAN, PHILIPPINES, POLAND, ROMANIA, SAUDI ARABIA, THAILAND, THE NETHERLANDS, UNITED KINGDOM.
Affected count
624
Manufactured in
7865 N 86th St, Milwaukee, WI, United States
Products
Mortara Surveyor S12 and Surveyor S19 Patient Monitors. Product Usage: They are indicated for use in adult & pediatric patient populations. They facilitate the monitoring of: Non-invasive blood pressure (NIBP), Impedance respiration, Invasive blood pressure (P1-P4), Temperature, Functional arterial oxygen saturation (SpO2), End-tidal & inspired CO2, ECG monitoring with arrhythmia and ST-segment analysis, 12-Lead resting ECG, Cardiac output (C)). The Mortara Surveyor Patient Monitor is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0110-2015

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief