FDA-Device2019-10-16Class II

Centurion CIRCLAMP W/1.3 CM BELL-single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 330CR

Centurion Medical Products Corporation
Hazard

A mismatch between the angle of the hole in the base and the angle of the bell, which may cause the clamp to remove foreskin without the use of a scalpel.

Sold states
Nationwide
Affected count
204 each
Manufactured in
100 Centurion Way, N/A, Williamston, MI, United States
Products
Centurion CIRCLAMP W/1.3 CM BELL-single-use, disposable instrument used to compress the foreskin of the penis during circumcision of a male infant. ReOrder: 330CR

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0121-2020

Don't want to check this manually?