FDA-Device2023-10-25Class II

Functional MR V1.0

OLEA MEDICAL
Hazard

When exporting regions of interest in DICOM SEGMENTATION format, when simultaneously exporting multiple volumes of interest, a functional magnetic resonance software bug may cause file contents and associated series descriptions to not match, which could lead to a misinterpretation of clinical data.

Sold states
US: MI, MA
Affected count
2
Manufactured in
93 AVANUE DU SORBIER,, ZONE ATHELIA IV, LA CIOTAT, N/A, France
Products
Functional MR V1.0

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0122-2024

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief