FDA-Device2023-10-25Class II
Functional MR V1.0
OLEA MEDICAL
Hazard
When exporting regions of interest in DICOM SEGMENTATION format, when simultaneously exporting multiple volumes of interest, a functional magnetic resonance software bug may cause file contents and associated series descriptions to not match, which could lead to a misinterpretation of clinical data.
Sold states
US: MI, MA
Affected count
2
Manufactured in
93 AVANUE DU SORBIER,, ZONE ATHELIA IV, LA CIOTAT, N/A, France
Products
Functional MR V1.0
Official notice