FDA-Device2021-10-27Class II

Atellica IM 1600 Analyzer, SMN 11066000

Siemens Healthcare Diagnostics, Inc
Hazard

Ancillary reagent packs used on Atellica IM 1300 and Atellica IM 1600 analyzers may overflow when pierced while on an analyzer located at an altitude of greater than 350m (1148ft) above sea level. As a result, customers at these altitudes may observe an increased number of Reagent Volume Check Errors.

Sold states
Domestic distribution to AZ CA CO GA ID KS MO MT NC NE NM NV OH OK OR PA TN TX UT VA WA WI WY. Foreign distribution worldwide.
Affected count
1733
Manufactured in
333 Coney St, N/A, East Walpole, MA, United States
Products
Atellica IM 1600 Analyzer, SMN 11066000

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0125-2022

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief