FDA-Device2016-10-26Class II

CORE¿ Suction Irrigation Handpiece with 5mm x 32cm Length Probe for Single Solution Bags, Catalog Number CD8185 The CORE Trumpet (CD81XX, CD83XX, and CD84XX) has applications in gynecological laparoscopy, laparoscopic cholecystectomy and other laparoscopic procedures.

ConMed Corporation
Hazard

The packaging seal may contain a crease. The crease may result in an open channel. These channels may compromise sterility of the product.

Sold states
Worldwide Distribution - US (Nationwide) Internationally: to Australia, Austria, Uruguay, Denmark, Germany, Ireland, Italy, United Kingdom, Guatemala, Czech Republic & Canada.
Affected count
N/A
Manufactured in
525 French Rd, N/A, Utica, NY, United States
Products
CORE¿ Suction Irrigation Handpiece with 5mm x 32cm Length Probe for Single Solution Bags, Catalog Number CD8185 The CORE Trumpet (CD81XX, CD83XX, and CD84XX) has applications in gynecological laparoscopy, laparoscopic cholecystectomy and other laparoscopic procedures.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0129-2017