FDA-Device2020-10-28Class II
Yuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Part 143301F0 (US) Part 143301B0 (OUS)
Maquet Cardiovascular Us Sales, Llc
Hazard
Plastic socket of the IEC 60320-1 socket pulled out of the metal cage and electronic parts exposed; where the socket was pulled out and main cable is still connected to the main socket (power source), there is a risk that a user or third person may suffer an electric shock when coming in contact with the exposed electronic components
Sold states
Worldwide distribution - US Nationwide and the countries of ARGENTINA, AUSTRALIA, BAHRAIN, BOLIVIA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, FINLAND, FRANCE GERMANY, HONG KONG, HUNGARY, INDIA, INDONESIA, IRAQ, ISRAEL, ITALY, JAPAN, KENYA, KOREA, KUWAIT, LATVIA, LEBANON, MACAU, MACEDONIA, MALAYSIA, MEXICO, NEW ZEALAND, OMAN, POLAND, QATAR, ROMANIA, RUSSIAN FEDERATION, RWANDA, SAUDI ARABIA, SERBIA, SINGAPORE, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TUNESIA, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM.
Affected count
62 units US
Manufactured in
45 Barbour Pond Dr, N/A, Wayne, NJ, United States
Products
Yuno Mobile Operating Table - Product Usage: Intended for support and position the patient immediately before, during and after surgical interventions as well as for examination and treatment. Part 143301F0 (US) Part 143301B0 (OUS)
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0132-2021Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief