FDA-Device2022-11-02Class III
LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Control Kit Catalog #: L016080109002 (EUA) The LumiraDx SARS-CoV-2 Quality Controls are intended for liquid quality control testing performed on the LumiraDx Instrument when used with the LumiraDx SARS-CoV-2 Ag Test Strip. The Quality Controls provide users with assurance that the device is performing within specification
LumiraDx
Hazard
Assigned vial lot numbers for the incorrect type of product. SARS-CoV-2 & Flu A/B lot numbers were incorrectly assigned to the positive and negative vial lot numbers for SARS-CoV-2 Ag LQCs within batch 2002204017, which should have been assigned SARS-CoV-2 Ag lot numbers resulting in an error message and preventing use of the LQC
Sold states
US Nationwide distribution.
Affected count
83 kits
Manufactured in
221 Crescent St, N/A, Waltham, MA, United States
Products
LumiraDx SARS-CoV-2 Antigen (Ag) Liquid Quality Control Kit Catalog #: L016080109002 (EUA) The LumiraDx SARS-CoV-2 Quality Controls are intended for liquid quality control testing performed on the LumiraDx Instrument when used with the LumiraDx SARS-CoV-2 Ag Test Strip. The Quality Controls provide users with assurance that the device is performing within specification
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0135-2023Don't want to check this manually?