FDA-Device2013-12-04Class II

Siemens RAD Fluoro Uro system (Luminos dRF and Uroskop Omnia with software version VD10A/G Image intensified Fluoroscopic X-ray system

Siemens Medical Solutions USA, Inc
Hazard

When using systems operating with software versions VD10A/G during a RAD examination, a malfunction can result in radiation continuing longer than required resulting in an overexposed image that is not of diagnostic quality, prompting repetition of the examination.

Sold states
Worldwide Distribution - USA (nationwide) and Internationally to Canada.
Affected count
59
Manufactured in
51 Valley Stream Pkwy, N/A, Malvern, PA, United States
Products
Siemens RAD Fluoro Uro system (Luminos dRF and Uroskop Omnia with software version VD10A/G Image intensified Fluoroscopic X-ray system

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0140-2014

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief