FDA-Device2024-11-06Class I

GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3) 2082844-002-01058424; 4) 2082844-002-01069017; 5) 2082844-002-01089666; 6) 2082844-002-01090474; 7) 2082844-002-01091269; 8) 2082844-002-01093489; 9) 2082844-002-01113262; 10) 2082844-002-01113711; 11) 2082844-002-01123085; 12) 2082844-002-01123918; 13) 2082844-002-01128966; 14) 2082844-002-01133042; 15) 2082844-002-01146447; 16)

DATEX--OHMEDA, INC.
Hazard

Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.

Sold states
worldwide
Affected count
224 units
Manufactured in
9900 W Innovation Dr, N/A, Wauwatosa, WI, United States
Products
GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3) 2082844-002-01058424; 4) 2082844-002-01069017; 5) 2082844-002-01089666; 6) 2082844-002-01090474; 7) 2082844-002-01091269; 8) 2082844-002-01093489; 9) 2082844-002-01113262; 10) 2082844-002-01113711; 11) 2082844-002-01123085; 12) 2082844-002-01123918; 13) 2082844-002-01128966; 14) 2082844-002-01133042; 15) 2082844-002-01146447; 16) 2082844-002-100205; 17) 2082844-002-100749; 18) 2082844-002-100752; 19) 2082844-002-100995; 20) 2082844-002-101425; 21) 2082844-002-523548; 22) 2082844-002-529928; 23) 2082844-002-594800; 24) 2082844-002-643544; 25) 2082844-002-925841; 26) 2082844-002-965540; 27) 2082844-002-982916; 28) 2082844-002-983420; 29) 2082844-002-993905; 30) 2082844-002-995018; 31) 2082844-002-995799; 32) 2082844-002-998943; Radiant infant warmer

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0141-2025