FDA-Device2022-11-09Class I
Hillrom Hospital Bed Accessory, Watch Care Incontinence Management System for Progressa Bed, Product Code/Part Number P00697903.
Hazard
Radio frequency emission from a functioning WatchCare device may potentially impact other devices (including, but not limited to telemetry devices, bladder scanner, fetal monitor/doppler, infusion pumps, and insulin pumps).
Sold states
US
Affected count
1137 units
Manufactured in
1 Baxter Pkwy, N/A, Deerfield, IL, United States
Products
Hillrom Hospital Bed Accessory, Watch Care Incontinence Management System for Progressa Bed, Product Code/Part Number P00697903.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0143-2023More Baxter Healthcare Corporation recalls
- FDA-Device2026-07-08Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device
- FDA-Device2026-07-08Adapter for Head Positioning, Model/Catalog Number: 1739994, when used in conjunction with the side rails of the Upper Back Section H V U component (product code 1909813) on Baxter operating tables
- FDA-Device2026-07-08Hillrom VOLARA System P.CIRCUIT 5KIT AC, (Single Patient Use Circuit), Model/Catalog Number REF M08474; Oscillation Lung Expansion Therapy, positive pressure breathing device
- FDA-Device2026-07-08Hillrom VOLARA System P.CIRCUIT KIT (Single Patient Use Circuit), Model/Catalog Number REF M08085; Oscillation Lung Expansion Therapy, positive pressure breathing device
- FDA-Device2026-07-08Hillrom VOLARA P.CIRCUIT 5KIT (System Single Patient Use Circuit), Model/Catalog Number REF M08270; Oscillation Lung Expansion Therapy, positive pressure breathing device
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief