FDA-Device2014-11-05Class II

Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.

Tecan US, Inc.
Hazard

Distortion (noise interference) of the liquid level capacitance can result in air aspiration instead of liquid aspiration. This might cause a false signal (air detection) with possible dispense errors resulting in erroneous results.

Sold states
US Distribution to IL.
Affected count
2 units
Manufactured in
9401 Globe Center Drive, Suite 140, Morrisville, NC, United States
Products
Freedom EVO-2 100 Base Unit, with Air LiHa Arms The Freedom EVO is an open automation platform product for general laboratory use. It is intended for routine laboratory tasks, such as general purpose pipetting and general purpose liquid handling and robotic processes.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0149-2015