FDA-Device2015-11-11Class I

HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31. Anesthesiology: The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients

Hamilton Medical, Inc.
Hazard

Ventilation and alarms of a HAMILTON-G5 ventilator can be suppressed unintentionally after the activation of a suctioning maneuver by the operator; this situation can occur regardless of the selected patient group (neonatal, pediatric and adult).

Sold states
Nationwide Distribution
Affected count
1128
Manufactured in
4990 Energy Way, N/A, Reno, NV, United States
Products
HAMILTON-G5 Ventilator with software versions between V2.00 and V2.31. Anesthesiology: The HAMILTON-G5 ventilator is designed for intensive care ventilation of adult and pediatric patients, and optionally infant and neonatal patients

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0151-2016

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief