FDA-Device2022-11-09Class II

THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electrophysiological mapping (stimulation and recording) and, in conjunction with an RF generator, for cardiac ablation.

Johnson & Johnson Surgical Vision, Inc.
Hazard

Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.

Sold states
Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO, MI, GA, OR, MO, SD, CO, OH, FL, NJ, TX, IL, WI and the countries of CROATIA, GERMANY, HUNGARY.
Affected count
91
Manufactured in
31 Technology Dr Ste 200, N/A, Irvine, CA, United States
Products
THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electrophysiological mapping (stimulation and recording) and, in conjunction with an RF generator, for cardiac ablation.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0169-2023