FDA-Device2025-10-22Class II
Basic Biopsy Tray; Item Number/Description: AKRB26D ROBOTIC, AMGL09AA GENERAL LAPAROSCOPY, ANLC35AD LAP CHOLE PACK - 205956, ANLP25AJ LAPROSCOPY PACK - 205958, FYLA38E LAPCHOLAPPY PACK, IHDA88AH DIVINCI GYN URO, IHLC11AE LAP GENERAL PACK, MHLP45AJ LAPAROSCOPY PACK - 206005, MHRB88AO ROBOTIC PACK - 242561, RCLC51C LAP CHOLE RCH 0246969, SMLC53C LAP CHOLE PACK, UDLC17AE DSC LAP CHOLE PACK - 206062, UIRU84AE ROBOTIC UROLOGY PACK (PS 060571)
American Contract Systems Inc.
Hazard
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.
Sold states
US Nationwide distribution in the states of SD, IA, MN, WA, IL.
Affected count
32,433 total
Manufactured in
1601 Se Gateway Dr Ste 120, Grimes, IA, United States
Products
Basic Biopsy Tray; Item Number/Description: AKRB26D ROBOTIC, AMGL09AA GENERAL LAPAROSCOPY, ANLC35AD LAP CHOLE PACK - 205956, ANLP25AJ LAPROSCOPY PACK - 205958, FYLA38E LAPCHOLAPPY PACK, IHDA88AH DIVINCI GYN URO, IHLC11AE LAP GENERAL PACK, MHLP45AJ LAPAROSCOPY PACK - 206005, MHRB88AO ROBOTIC PACK - 242561, RCLC51C LAP CHOLE RCH 0246969, SMLC53C LAP CHOLE PACK, UDLC17AE DSC LAP CHOLE PACK - 206062, UIRU84AE ROBOTIC UROLOGY PACK (PS 060571)
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0171-2026Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief