FDA-Device2015-11-04Class II

ADVIA Chemistry XPT Chemistry Systems, Siemens Material Number (SMN) 10723034. An automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes.

Siemens Healthcare Diagnostics, Inc.
Hazard

ISE module may produce discrepant, unflagged results for Sodium, Potassium or Chloride when the respective electrode is left unplugged after replacement, washing or maintenance activities. The calibration and 2-level QC will fail if the electrode is left unplugged. The IFU states to calibrate the ISEs after replacement and to run two levels of controls before running patient samples.

Sold states
Worldwide Distribution -- US, Australia, Bahrain, Denmark, Egypt, Egypt, Finland, France, Germany, Italy, Netherlands, New Zealand, Norway, Republic Korea, Singapore, Slovakia, Spain, and the United Kingdom.
Affected count
Domestic: 15 units; Foreign: 59 units
Manufactured in
511 Benedict Ave, N/A, Tarrytown, NY, United States
Products
ADVIA Chemistry XPT Chemistry Systems, Siemens Material Number (SMN) 10723034. An automated, clinical chemistry analyzer that runs tests on serum, plasma, urine, or cerebral spinal fluid in random access and batch modes.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0183-2016