FDA-Device2014-11-19Class II

GE Healthcare Infinia (including Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye, Infinia II Hawkeye 4, Infinia II VC, Infinia II VC Hawkeye, and Infinia II VC Hawkeye 4) Models: H3000WC, H3000WD, H3000WE, H3000WM, H3000WN, H3000WP, H3000WR, H3000WT, H3000WW, H3000WY, H3000WZ, H3000YF, H3000YG, H3000YM, H3000YS, H2400YJ, H2400YK. Hawkeye 4 Option For Dual-Head Variable Angle Gamma Camera, Hawkeye Option for Dual-Head Variable Angle Gamma Camera (this includes GE Quasar Nuclear Me

GE Healthcare, LLC
Hazard

GE is issuing this recall due to a potential safety issue related to a portion of the system falling onto the patient during a scan due to fasteners being loose that secured the camera to the gantry. GE is updating the Preventative Maintenance procedure and schedule, and added an additional Preventative Maintenance check for loose fasteners on the Nuclear Medicaine systems.

Sold states
Worldwide Distribution: US (nationwide) including Guam, Puerto Rico and DC; and countries of: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BELGIUM, BELRARUS, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, ECUADOR, EGYPT, ESTONIA, FINLAND, FRANCE, GEORGIA, GERMANY, HUNGARY, GREECE, GUADELOUPE, HONG KONG, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, REPUBLIC OF KOREA, KUWAIT, LATVIA, LEBANON, LITHUANIA, MALAYSIA, MALTA,MARTINIQUE, MEXICO, MOROCCO, NAMIBIA, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL QATAR, ROMANIA, RUSSIA, SAUDI ARABIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SPAIN, SWEDEN,SWITZERLAND, TAIWAN, THAILAND, TUNISIA, TURKEY, UNITED KINGDOM, UKRAINE, URUGUAY, VENEZUELA, VIET NAAM.
Affected count
2479
Manufactured in
3000 N Grandview Blvd, Waukesha, WI, United States
Products
GE Healthcare Infinia (including Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye, Infinia II Hawkeye 4, Infinia II VC, Infinia II VC Hawkeye, and Infinia II VC Hawkeye 4) Models: H3000WC, H3000WD, H3000WE, H3000WM, H3000WN, H3000WP, H3000WR, H3000WT, H3000WW, H3000WY, H3000WZ, H3000YF, H3000YG, H3000YM, H3000YS, H2400YJ, H2400YK. Hawkeye 4 Option For Dual-Head Variable Angle Gamma Camera, Hawkeye Option for Dual-Head Variable Angle Gamma Camera (this includes GE Quasar Nuclear Medicine Systems). Nuclear Medicine Systems Infinia devices Product Usage: The intended use of the Quasar system is to perform general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body, using a variety of scanning modes supported by various acquisition types and optional imaging features designed to enhance image quality in Oncology, Cardiology, Neurology and other clinical diagnostic imaging applications.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0184-2015