FDA-Device2021-11-10Class II

FIVE S 3.5x65, Flexible Intubation Videoscope, REF: 091361-01 (1 Unit), and 091361-06 (6 Units)

Karl Storz Endoscopy
Hazard

During a regular requalification review, affected flexible intubation video endoscopes did not achieve the required sterility assurance level. There is a risk that the patient may be exposed to a higher risk of infection.

Sold states
US: SC, NH, UT, LA, FL, TX, MA, MO, VA, IA, CA, IL, NC, VT, AL, GA, PA, TN, WI, NY, CO, AZ, NV, ME, MS, DC, IN, KY, MD, OH, MI, WA, KS, OR, AR, MN, NJ, NM, WV, CT. OUS: Canada
Affected count
532
Manufactured in
2151 E Grand Ave, N/A, El Segundo, CA, United States
Products
FIVE S 3.5x65, Flexible Intubation Videoscope, REF: 091361-01 (1 Unit), and 091361-06 (6 Units)

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0184-2022