FDA-Device2020-10-28Class II

5 IN(13cm) APPX 0.87ml, EXT SET,NanoClave T-CONNECTOR; 1 unit per pouch, 50 pouches per case; UDI: (01)10887709052809(17)250101(30)50(10)4599980.

ICU Medical, Inc.
Hazard

Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.

Sold states
Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
Affected count
Total of all products (Listed #1 thru 101) = 304735 units
Manufactured in
951 Calle Amanecer, N/A, San Clemente, CA, United States
Products
5 IN(13cm) APPX 0.87ml, EXT SET,NanoClave T-CONNECTOR; 1 unit per pouch, 50 pouches per case; UDI: (01)10887709052809(17)250101(30)50(10)4599980.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0185-2021