FDA-Device2018-10-31Class I
CyPass System 241, Intraocular Pressure Lowering Implant, CYPASS MICRO-STENT, CYPASS APPLIER, REF: 8065754001. For ophthalmic use.
Alcon Research, LTD.
Hazard
Analysis of the completed dataset from the COMPASS-XT long-term study showed that the CyPass Micro-Stent group experienced statistically significant endothelial cell loss (ECL) compared to the group who underwent cataract surgery alone.
Sold states
Worldwide distribution. US nationwide including Puerto Rico, Argentina, Australia, Austria, Canada, Chile, Colombia, Costa Rica, Denmark, Dominican Republic, Ecuador, France, Finland, Germany, Hong Kong, Iceland, Italy, Mexico, Peru, Singapore, Spain, Sweden, Switzerland, United Kingdom, and Uruguay.
Affected count
14264 units
Manufactured in
714 Columbia Ave, N/A, Sinking Spring, PA, United States
Products
CyPass System 241, Intraocular Pressure Lowering Implant, CYPASS MICRO-STENT, CYPASS APPLIER, REF: 8065754001. For ophthalmic use.
Official notice