FDA-Device2024-10-30Class II
medex Arterial Kit 1/EA, Product Code REF MX7827
Smiths Medical ASD, Inc.
Hazard
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Sold states
Worldwide distribution.
Affected count
2232 units
Manufactured in
6000 Nathan Ln N, N/A, Minneapolis, MN, United States
Products
medex Arterial Kit 1/EA, Product Code REF MX7827
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-0187-2025Don't want to check this manually?